Original Article


Evaluation of body fluid mode of Sysmex XN-9000 for white blood cell counts in cerebrospinal fluid

Vincenzo Roccaforte, Massimo Daves, Vanessa Proserpio, Flavia Sciarini, Rosella Sangiorgio, Arcangelo Costanzo, Angela M. Di Pierro, Salvadori Ugo, Giuseppe Lippi, Claudio Bonato

Abstract

Background: This study was planned to evaluate the analytical performance of the novel and fully automated Sysmex XN-9000 analyzer for rapid analysis of cerebrospinal fluid (CSF) samples.
Methods: Forty-four CSF samples were used for method comparison studies between Sysmex XN-9000 body fluid mode and conventional optical microscopy. The bias between data obtained with the two methods was estimated with Bland-Altman plot analysis. The analytical evaluation also included the assessment of imprecision, linearity and carry-over.
Results: A good agreement was found between results obtained with Sysmex XN-9000 body fluid mode and optical microscopy. The mean bias was 1.6×106 cells/L for total white blood cells (95% CI: −21.8×106 to 25.1×106 cells/L), 1.3×106 cells/L for polymorphonuclear cells (95% CI: −13.9×106 to 16.5×106 cells/L) and −0.6×106 cells/L for mononuclear cells (95% CI: −21.5×106 to 20.3×106 cells/L). The carryover was found to be lower than 0.01% and the imprecision was lower than 5%. The XN-9000 body fluid mode was also characterized by excellent linearity in the range of values comprised between 85×106–3,197×106 cells/L, with correlation coefficients (r) always equal to 1.00 (P<0.001).
Conclusions: The Sysmex XN-9000 body fluid mode displays excellent analytical performance in terms of imprecision, linearity, carry-over and comparability with conventional optical microscopy, so that it may be used as a first-line, screening technique for rapid analysis of CSF samples referred for both routine and, especially, for urgent testing.

Download Citation